Unimet 810ST CH Ausführung
Standard-compliant storage of the initial measurement value as a reference
Tests in accordance with IEC 62353 and IEC 60601-1
Customizable test sequences
Data storage for over 10,000 data records
The only safety tester that stores the “initially measured value” in accordance with standards, so it can be used as a reference value for later comparisons.
The UNIMET 810ST allows you to customize all tests yourself. Test report history (test log) is stored in the included PC software.
Features and Functions:
· Easy to use and operate thanks to the Windows interface
· Data exchange and storage via Control Center
· Automatic, semi-automatic, or manual test sequence
· Data entry via touchscreen, keyboard, or barcode
· Customizable visual, electrical, and functional tests
· Design your own test sequences
· Data storage > 10,000 data records
· Filter function for quick data retrieval
· Test schedule management
· Multi-client capability
· Catalog systems
· Test probe with two switching contacts—for semi-automatic testing of parts not connected to PE
· Compatible with common application programs, e.g., visual FM, MT Data, and Fundamed
· Dimensions: (WxHxD) approx. 300x277x126 mm
· Weight: approx. 3.5 kg
Tests in accordance with:
· IEC 60601-1, 3rd Edition 1)
· IEC 62353:2014
· DIN EN 62353 (VDE 0751-1):2015-10
· IEC 61010-1:2011-07 1)
· ÖVE/ÖNORM EN 62353:2009-01
· DIN VDE 0701-0702:2008-06
· ÖVE E8701-1:03-01
1) These standards can be activated as a license in the base unit.
Standard-compliant measurements
· Protective conductor resistance
· Insulation resistance
· Earth leakage current
· Protective conductor current
· Equipment leakage current
· Touch current
· Touch voltage
· Direct measurement
· Residual current measurement
· Equivalent leakage current measurement
· Equivalent patient leakage current measurement
· AC and DC patient leakage current
· AC and DC patient auxiliary current
Interfaces:
· RS-232 port
· USB port for printer, scanner, keyboard, and USB flash drive
· PS/2 port for scanners and keyboards
· Network connection via Ethernet (optional)
Standard accessories:
· Sturdy protective and carrying case with a universal carrying design
· Test lead with test probe
· Test lead with test clamp
· Null modem cable
· VK 701-7 for testing low-voltage equipment
· UNIMET Control Center data management software
Service:
· Calibration interval: 36 months
· Warranty period: 36
· monthsFree hotline
· Future-proof thanks to software updates
· Optional training seminars
The software license for IEC 60601-1 must be ordered separately.