isoMED427PCH-2 Isolations- und Trafoüberwachung
Insulation Monitoring for Medical IT Systems
Load and temperature monitoring of the transformer
Expandable for insulation fault detection with EDS
Compliant with IEC 60364-7-710 and IEC 61557-8
The ISOMETER isoMED427P monitors the insulation resistance of ungrounded AC circuits (medical IT systems). At the same time, it monitors the IT system transformer for load current and overtemperature. When used in conjunction with insulation fault locators from the EDS… series and the corresponding current transformers, the isoMED427P can be used to set up a system for locating insulation faults.
Features:
· Insulation monitoring device for medical IT systems
· Load and temperature monitoring for IT system transformers
· Test current generator for insulation fault detection equipment
· Can be expanded into an insulation fault detection system
Features:
· Insulation monitoring for medical IT systems
· Adjustable threshold value for insulation monitoring
· Test current generator for insulation fault detection systems
· Load and temperature monitoring for IT system transformers
· Adjustable threshold for load current
· Temperature monitoring with a PTC thermistor or bimetallic switch
· Self-monitoring with automatic notification
· PE connection monitoring
· Internal and external test buttons
· Indicator LEDs for operation, alarm 1, alarm 2
· Programmable alarm relay: selectable operating/quiescent current
· Compact 2-module housing (36 mm)
· BMS interface
Applications:
Medical IT systems in accordance with IEC 60364-7-710, IEC 61557-8, IEC 61557-9, and DIN VDE 0100-710